Agilent InfinityLab LC Series Vialsamplers User Manual 246
8Maintenance
Exchange the Rotor Seal
1 Open the doors of the module. 2 Remove all capillaries from the injection valve with a
1/4 inch wrench.
3 Loosen each fixing bolt two turns at a time. Remove the
bolts from the head.
4 Remove the stator head and stator ring.