Agilent InfinityLab LC Series Vialsamplers User Manual 268
8Maintenance
Exchange the Wash Port Assembly
21 Install the safety cover (1.) and fix the screw (2.). 22 In Lab Advisor, select Next to move the needle into the
needle seat and then Back to finish the maintenance
procedure.
OR
In the Local Controller, select Continue to move the nee-
dle into the needle seat and then Exit to finish the main-
tenance procedure.
23 Select Back to finish the procedure.
24 Close the doors.
Next Steps:
25 Purge the wash port. Check if solvent is delivered from
the peristaltic pump.
26 Perform a System Pressure Test.
(1.)
(2.)