Agilent InfinityLab LC Series Vialsamplers User Manual 271
8Maintenance
Replace the Sample Cooler/Sample Thermostat
WARNING
In the event of a damage
Keep open fire or sources of ignition away from the device.
Ventilate the room for several minutes.
Do not use the Sample Thermostat any more.
WARNING
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy
objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
CAUTION
Routing of the condensation tubing
Proper routing of the condensation tubing is critical for correct condensate
drainage.
Do not place the sampler directly on the bench.
CAUTION
Condensate inside the cooler or thermostat
Damage to the electronics
Unplug the power cords.
Drain off all condensate before dismounting the sample cooler or
thermostat.
Make sure that there is no condensate left.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals