Agilent InfinityLab LC Series Vialsamplers User Manual 46
2 Site Requirements and Specifications
Performance Specifications
Minimum sample vol-
ume
1 µL from 5 µL sample in 100 µL
microvial, or 1
µL from 10 µL
sample in 300
µL microvial.
Needle height offset has to be
adapted to ensure that needle
doesn’t touch vial bottom.
Default needle height = 0
equates to 2
mm above the vial
bottom.
Instrument Control LC & CE Drivers A.02.17 or
above
Instrument Control Framework
(ICF) A.02.05 or above
Instant Pilot (G4208A) with firm-
ware B.02.22 or above
InfinityLab LC Companion
(G7108A)
Lab Advisor B.02.10 or above
For details about supported
software versions refer to the
compatibility matrix of your ver
-
sion of the LC & CE Drivers
Communication Controller Area Network (CAN),
Local Area Network (LAN),
ERI: ready, start, stop and
shut-down signals
Maintenance and
safety-related features
Extensive diagnostics, error
detection and display with Agi
-
lent Lab Advisor software
Leak detection, safe leak han-
dling, leak output signal for
shutdown of pumping system,
and low voltages in major main
-
tenance areas
GLP features Early maintenance feedback
(EMF) for continuous tracking
of instrument usage with
user-settable limits and feed
-
back messages. Electronic
records of maintenance and
errors.
Housing All materials recyclable.
Table 4 Performance Specifications G7129C
Type Specification Comments Method/Conditions