Agilent InfinityLab LC Series Vialsamplers User Manual 78
3Using the Module
Install the Optional Integrated Column Compartment
3 Push the latch of the Integrated Column Compartment
to the left.
4 Open the flap door.
5 Connect the injector valve (port 6) of the Vialsampler to
the inlet port of the ICC heat exchanger with a suitable
capillary (see
“Capillary Connections” on page 278 for
recommendations).
6 Connect a suitable capillary to the respective outlet port
of the ICC heat exchanger (see
“Capillary
Connections” on page 278 for recommendations).