Agilent InfinityLab LC Series Vialsamplers User Manual 83
3Using the Module
Install the Optional Integrated Column Compartment
3 Connect the injector valve (port 6) of the Vialsampler to
the inlet port of the ICC heat exchanger with a suitable
capillary (see
“Capillary Connections” on page 278 for
recommendations).
4 Connect a suitable capillary to the respective outlet port
of the ICC heat exchanger (see
“Capillary
Connections” on page 278 for recommendations).
5 Tighten the capillary fitting to make the connection
pressure tight.
Do not overtighten the fitting (maximum torque 2.5 Nm).
The use of capillaries with SL or SX fittings are recom-
mended for a better handling.
6 Mount the column holders on the heat exchanger.