EasyManuals Logo
Home>Agilent Technologies>Water Pump>MSD 5977B Series

Agilent Technologies MSD 5977B Series User Manual

Agilent Technologies MSD 5977B Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
5 Operating in EI Mode
To Set the GC/MSD Interface Temperature from MassHunter
124 5977B Series MSD Operating Manual
To Set the GC/MSD Interface Temperature
from MassHunter
Procedure
1 From Instrument Control view select Instrument>GC Edit Parameters.
2 Click the Aux Heaters icon to edit the interface temperature. (See Figure 38.)
3 Select On to turn on the heater and type the setpoint in the Value °C column.
The typical setpoint is 280 °C.
4 Click Apply to download setpoints, or click OK to download setpoints and
close the window.
5 To make the new settings part of the method, select Save from the Method menu.
Figure 38 8890 GC edit parameters window
CAUTION
Ensure that the carrier gas is turned on and the column has been purged of air
before heating the GC/MSD interface or the GC oven.
When setting the GC/MSD interface temperature, never exceed the maximum
for your column.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSD 5977B Series and is the answer not in the manual?

Agilent Technologies MSD 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSD 5977B Series
CategoryWater Pump
LanguageEnglish

Related product manuals