4 Installing 7890 GC Columns
To Calibrate Column Flow Linear Velocity in a 7890 GC
5977B Series MSD Operating Manual 95
To Calibrate Column Flow Linear Velocity in a
7890 GC
The steps listed below apply to 7890 and 6890 series GCs only. Capillary
columns must be calibrated prior to use with the MSD.
Procedure
1 Set Data Acquisition for splitless manual injection, and set up a real time plot
to monitor m/z 28.
2 Press [Prep Run] on the GC keypad.
3 Inject 1 µL of air into the GC inlet and press [Start Run].
4 Wait until a peak elutes at m/z 28. Note the retention time.
5 In the Instrument Control view, select GC Parameters from the Instrument
menu.
6 Select the Configuration tab and then select the Columns tab.
7 Select your installed column from the table.
8 Click Calibrate to display the Calibrate Column dialog.
9 Click Calc Length in the If unretained peak holdup time is known section to
display the Calculate Column Length dialog. (See Figure 25.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSD 5977B Series and is the answer not in the manual?

Agilent Technologies MSD 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSD 5977B Series
CategoryWater Pump
LanguageEnglish

Related product manuals