7 General Maintenance
To Clean an EI XTR, SS, or Inert Source
204 5977B Series MSD Operating Manual
To Clean an EI XTR, SS, or Inert Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. (See “To Disassemble the EI SS or EI Inert
Source” on page 198 or “To Disassemble the EI XTR Source” on page 201.)
2 Collect the following parts to be cleaned for an EI SS or Inert source: (See
Figure 66 on page 206.)
Repeller
Interface socket
Source body
Repeller block insert
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSD 5977B Series and is the answer not in the manual?

Agilent Technologies MSD 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSD 5977B Series
CategoryWater Pump
LanguageEnglish

Related product manuals