EasyManuals Logo
Home>Agilent Technologies>Water Pump>MSD 5977B Series

Agilent Technologies MSD 5977B Series User Manual

Agilent Technologies MSD 5977B Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
6 Operating in CI Mode
General Guidelines
144 5977B Series MSD Operating Manual
General Guidelines
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
Always verify the MSD is performing well in EI mode before switching to CI.
(See “To Verify EI System Performance” on page 131.)
Ensure the CI source and GC/MSD interface tip seal are installed.
Ensure the reagent gas plumbing has no air leaks. This is determined in PCI
mode, checking for m/z 32 after the methane pretune.
Ensure the reagent gas inlet line is equipped with gas purifiers (not applicable
for ammonia).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSD 5977B Series and is the answer not in the manual?

Agilent Technologies MSD 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSD 5977B Series
CategoryWater Pump
LanguageEnglish

Related product manuals