3 Installing Intuvo 9000 GC Columns
Columns
56 5977B Series MSD Operating Manual
Columns
Many types of Intuvo 9000 GC columns can be used with the MSD, but there are
some restrictions.
During tuning or data acquisition, the rate of column flow into the MSD should
not exceed the maximum recommended flow. Therefore, there are limits to
column length and flow. Exceeding recommended flow will result in degradation
of mass spectral and sensitivity performance.
Remember that column flows vary greatly with temperature, requiring
measurements of the actual flow to be taken. Use Table 6 to determine an
acceptable column flow.
Table 6 Gas flows
Feature Gas flows
High vacuum pump Diffusion Turbo
Optimal He column flow mL/min 1 1 to 2
Maximum recommended gas flow mL/min
*
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H
2
flow (if
applicable). Based on helium gas use. For other gases the maximum flow will vary.
1.5 4
Maximum gas flow, mL/min
†
† Expect degradation of spectral performance and sensitivity.
26.5
Max column id 0.53 mm
(30 m)
0.53 mm
(30 m)
JetClean option H
2
flow NA 0.4 mL/min