1Introduction
The oil pan under the foreline pump can be a fire hazard
5977B Series MSD Operating Manual 23
The oil pan under the foreline pump can be a
fire hazard
Oily rags, paper towels, and similar absorbents in the oil pan could ignite and
damage the pump and other parts of the MSD.
WARNING
The insulation around the GC inlets, detectors, valve box, and the insulation
cups is made of refractory ceramic fibers. To avoid inhaling fiber particles, we
recommend the following safety procedures:
Ventilate your work area
Wear long sleeves, gloves, safety glasses, and a disposable dust/mist
respirator
Dispose of the insulation in a sealed plastic bag
Wash your hands with mild soap and cold water after handling the
insulation
WARNING
Combustible materials (or flammable/nonflammable wicking material) placed
under, over, or around the foreline (roughing) pump constitutes a fire hazard.
Keep the pan clean, and do not leave absorbent material such as paper towels
in it.