EasyManuals Logo
Home>Agilent Technologies>Water Pump>MSD 5977B Series

Agilent Technologies MSD 5977B Series User Manual

Agilent Technologies MSD 5977B Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #238 background imageLoading...
Page #238 background image
8CI Maintenance
To Disassemble the CI Source
238 5977B Series MSD Operating Manual
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view Figure 78 and CI source parts list Table 24 on
page 239 while using this procedure.
1 Remove the CI source. (See“To Remove the CI Source” on page 235.)
2 Remove the filaments. (See“To Remove the CI Source Filament” on
page 247.)
3 Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts.
4 Disassemble the repeller assembly by removing the ceramic insulator from
the repeller.
5 Remove the setscrew securing the lenses to the source body.
6 Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSD 5977B Series and is the answer not in the manual?

Agilent Technologies MSD 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSD 5977B Series
CategoryWater Pump
LanguageEnglish

Related product manuals