7 General Maintenance
To Clean an EI XTR, SS, or Inert Source
5977B Series MSD Operating Manual 205
3 Collect the following parts to be cleaned for an EI XTR source: (See Figure 66
on page 206.)
Repeller
Repeller block insert
Source body
Extractor lens
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
CAUTION
If insulators are dirty, clean them with a cotton swab dampened with
reagent-grade methanol. If that does not clean the insulators, replace them. Do
not abrasively or ultrasonically clean the insulators.