EasyManua.ls Logo

Agilent Technologies MSD 5977B Series

Agilent Technologies MSD 5977B Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Operating in EI Mode
To Perform High-Mass Testing
134 5977B Series MSD Operating Manual
Results
Results will indicate the recommended amount to adjust AMU offset for high
mass. If your results are within 5 units of the targeted amount, there is no need to
make adjustments.
Adjustments
1 Verify the autotune file (atune.u, HES_atune.u, etune.u) has been loaded.
2 In Instrument Control view select Edit Tune Parameters from the Instrument
menu to display the Manual Tune dialog.
3 Click the Dynamic tab and then click the Amu Offset sub-tab.
4 Select the Enable This Lens checkbox.
5 Enter the recommend dynamic offset Voltage (V), and click OK.
6 Click Save to save this dynamic Amu Offset for the high-mass.
You can overwrite the existing autotune file to include high-mass
adjustment or save this file to a new name, for example, ATUNEHIGH.U.
When an autotune is performed, it will overwrite this dynamic Amu Offset
that was entered. This is why you might want to rename the tune.
7 Click Done to close the Manual Tune dialog.
8 Load the PFHT.M, then load the saved tune file, and then save the method.
9 Rerun test mixture (repeat high-mass checkout). If the correction is within
5 units, no further adjustments are required.

Table of Contents

Related product manuals