3 Installing Intuvo 9000 GC Columns
To Condition an Intuvo Capillary Column
5977B Series MSD Operating Manual 73
6 Program the column temperature to increase from 120 °C to the maximum
temperature limit for the column at a rate of 10 to 15 °C/min.
7 Hold at the maximum temperature for 30 minutes.
The column is conditioned and ready to be used with your method.
CAUTION
Never exceed the maximum column temperature, in the GC/MSD interface,
the GC oven, or the inlet.