EasyManua.ls Logo

Agilent Technologies MSD 5977B Series

Agilent Technologies MSD 5977B Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Installing 7890 GC Columns
To Calibrate Column Flow Linear Velocity in a 7890 GC
5977B Series MSD Operating Manual 97
Calculation for average linear velocity
Average linear velocity (cm/s) =
where:
L = Length of the column in meters
t = Retention time in seconds
Calculation for volumetric flow rate
Volumetric flow rate (mL/min) =
where:
D = Internal column diameter in millimeters
L = Column length in meters
t = Retention time in minutes
100 L
t
---------------
0.785 D
2
L
t
---------------------------

Table of Contents