8-10 Service Guide N5230-90024
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers
Error Terms N5230C
Error Term Data
The error term descriptions in this section include the following information:
• a table of the error terms
• description and significance of each error term
• measurements affected by each error term
• typical cause of failure for each error term
The same description applies to both the forward (F) and reverse (R) terms.
IMPORTANT Data are listed here as a convenience only. Detailed instrument specifications are listed in
the embedded help in the network analyzer.
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system verification. Refer to Chapter 3,
“Tests and Adjustments”.
Tabl e 8-1 Er ror Term D ata
a
a. The data in this table are uncorrected system performance. The values apply over an environmental temperature range
of 25 °C ±5 °C, with less than 1 °C deviation from the calibration temperature
.
Parameter
Frequency Range
10 MHz
to
45 MHz
b
b. All values for 10 MHz to 45 MHz are typical.
45 MHz
to
500 MHz
500 MHz
to
2 GHz
2 GHz
to
8 GHz
8 GHz
to
13.5 GHz
13.5 GHz
to
20 GHz
20 GHz
to
40 GHz
40 GHz
to
45GHz
45 GHz
to
50 GHz
Directivity
20 GHz models
40 GHz models
50 GHz models
>24dB
>22dB
>22dB
>24dB
>23dB
>23dB
>27dB
>23dB
>23dB
>21dB
>21dB
>21dB
>16dB
>16dB
>16dB
>16dB
>16dB
>16dB
>15dB
>15dB >15dB >13dB
Source Match
20 GHz models
40 GHz models
50 GHz models
>12dB
>12dB
>12dB
>20dB
>17dB
>17dB
>17dB
>17dB
>17dB
>12dB
>12dB
>12dB
>11dB
>11dB
>11dB
>10dB
>11dB
>11dB
>7dB
>7dB >7dB >7dB
Load Match
20 GHz models
40 GHz models
50 GHz models
>15dB
>12dB
>12dB
>22dB
>18dB
>18dB
>20dB
>18dB
>18dB
>12dB
>15dB
>15dB
>10dB
>12dB
>12dB
>9dB
>12dB
>12dB
>10dB
>10dB >5dB >5dB
Crosstalk
c
20 GHz models
40 GHz models
50 GHz models
c. All crosstalk values are typical.
88dB
>60dB
>60dB
>95dB
>60dB
>60dB
>96dB
>85dB
>85dB
>110dB
>85dB
>85dB
>116dB
>82dB
>82dB
>115dB
>82dB
>82dB
>72dB