The front panel assembly consists of the A1 keypad, the A2 display assembly, the A3 front panel interface
board, the touchscreen and touchscreen controller board, the inverter board, and the USB board. The
following tests verify the operation of the front panel assembly when the analyzer is in the measurement
mode. If the instrument fails to power up correctly, or it is difficult to verify due to a faulty display, refer to
“Power Up Troubleshooting” on page 4-6.
Refer to the following sections to verify the operation of the noted assemblies.
“Front Panel Keypad and RPG Test” on page 4-13
A2 Display Test” on page 4-15
“Checking the USB Board” on page 4-16
A3 Front Panel Interface Board” on page 4-16
If all assemblies are working correctly, continue troubleshooting with “Rear Panel Troubleshooting” on
page 4-17.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5230C and is the answer not in the manual?

Agilent Technologies N5230C Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5230C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals