Service Guide N5230-90024 8-17
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures
N5230C Option Enable Utility
Installing or Changing a Serial Number
It is necessary to reinstall the instrument serial number if the test set motherboard is replaced.
IMPORTANT Use extreme care when entering the serial number, as only one attempt is allowed.
1. To change a serial number, select
Change Serial from the Option Enable menu bar (see Figure 8-5 on
page 8-14). The current serial number is displayed. If no serial number has previously been entered, the
word “NONE” will be displayed.
2. VERY CAREFULLY, type the new serial number into the space provided and then click
Change.
3. If an error is made in entering the serial number, obtain a clear code from Agilent, enter the clear code in
the space provided, and then click
Clear. The correct serial number can then be entered.
NOTE To change an incorrect serial number, a clear-code password is required. Contact Agilent to
obtain the clear-code. Refer to “Contacting Agilent” on page 2-10.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5230C and is the answer not in the manual?

Agilent Technologies N5230C Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5230C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals