3-18 Service Guide N5230-90024
Tests and Adjustments PNA Series Microwave Network Analyzers
Preliminary Checks N5230C
Figure 3-7 Typical Cable Connector Repeatability Response
If the Cable Connector Repeatability Check Fails
1. Clean the cable and devices, and torque to specification. Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.