System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide 119
CAUTION
The ends of the sheath fluid in-line filter are designed to be different on
purpose to ensure the in-line filter is connected into the system in the
correct direction. Incorrect direction would lead to system malfunction.
CAUTION
When replacing the sheath fluid in-line filter, do not twist or kink the tubing.
Replace the Waste Filter
1 Make sure the instrument is in Ready or shut down status.
2 Open the front panel of the instrument. Locate the waste filter as shown
below.
Figure 98. Location of Waste Filter
3 Unscrew the Luer connectors at two ends of the waste filter to disconnect the
tubings from the filter.
CAUTION
Use lint free wipes to gently wipe of the sheath fluid from the tubing and
avoid sheath fluid dripping onto the instrument.
4 Replace the old filter with a new one and correctly connect the tubing to the
filter. Ensure the side of the filter labeled with "Inlet" is connected to the inlet
tubing as shown below.