System Maintenance
120 NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 99. Connect the Waste Filter
CAUTION
Finger tightening the Luer connectors only to avoid damage of the filters
and filter leakage.
5 Close the front panel. If the instrument is in shut down status, turn on the
instrument, and automatic start-up cleaning process will be initiated. Observe
the sheath fluid flows through the filter properly and make sure there is no
leakage through the filter-tubing connections. Then run the Priming process
when start-up process is completed and instrument is ready.
CAUTION
When replacing the waste filter, do not twist or kink the tubing.
WARNING
All biological samples or materials can contain infectious deadly disease.
Please be careful and always wear protective clothing, masks, and gloves
when handling the waste.
Replace NovoFlow Sheath Fluid Inlet Filter
NovoFlow sheath fluid inlet filter is placed inside the NovoFlow container to filter
out micro-particles or impurities in the sheath fluid.
1 Make sure the instrument is in Ready or shut down status.