System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide 123
Figure 102. Location of the 0.05 µm Sheath Fluid Ultrafiltration (UF) Filter rack
CAUTION
Use lint free wipes to gently wipe of the sheath fluid from the tubing and
avoid sheath fluid dripping onto the instrument.
4 Connect the outlet of the new filter to the female Luer connector, and the inlet
of the new filter to the male Luer connector. Secure the filter onto the filter
rack by gently sliding it through the open slit on the rack.
CAUTION
Ensure the filter is vertically mounted and all the tubing is not twisted.
NOTE
Under some situation when the Ultrafiltration Filter needs to be removed after initial
installation, disconnect the filter from the Luer connectors and connect the male
Luer connector directly to the female one.