EasyManuals Logo

Agilent Technologies NovoCyte Quanteon User Manual

Agilent Technologies NovoCyte Quanteon
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
System Maintenance
124 NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 103. Properly Installed 0.05 µm Sheath Fluid Ultrafiltration (UF) Filter
5 Open NovoExpress software. Click Instrument > Fluidic Maintenance >
Priming to prime the fluidic system. Ensure no fluidic leakage from the filter’s
inlet/outlet.
6 Load a sample tube with 1 mL Milli-Q or 0.2 µm filtered DI water onto the tube
holder. Create a new experiment. Set Stop Condition as 500 μL and Flow Rate
as 120 μL/min. Click the Run button in Experiment Control panel to start the
sample acquisition. This step is to flush sample tubing.
7 Close the front panel. The installation is completed, and the instrument is
ready to use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals