Troubleshooting
NovoCyte Quanteon Flow Cytometer Operator's Guide 135
not connected even
though the
instrument is
powered up
communication between the
instrument and workstation
Improper connection between
the instrument and
workstation
Reconnect the USB cable between the
instrument and the workstation.
More than one NovoExpress
instances are running
Only run one instance of NovoExpress software.
Unstable Signal or
gap in the data plot
Air bubbles exist in the flow
cell
Execute the Debubble function from
NovoExpress Software.
Sample tube does
not fit the sample
tube holder
Wrong type of sample tube is
used
Use 12 × 75 mm tube or 1.5 mL tube 2.0 mL
tube.
Sample Injection
Probe is not
properly aligned
with the wells or
tubes
Plate or tube rack is not
loaded correctly.
Load the plate or tube rack with the correct
orientation. Make sure the plate or tube rack is
fully seated onto the sample tray properly.
NovoSampler (or
NovoSampler Pro or
NovoSampler Q) is not
calibrated.
Re-calibrate the NovoSampler (or NovoSampler
Pro or NovoSampler Q). Refer to associated
operator's guide for more details.
Leakage from the
SIP
Waste tubing is deformed
Reconnect the waste tubing to ensure it is
properly connected.
Move the instrument forward as needed to
ensure sufficient space between the
instrument and wall. Refer to associated
instrument operator's guide for more
information on properly installing
instrument.
Waste filter is clogged
Replace the waste filter following the
procedures described in associated instrument
operator's guide.
Leakage from
Sampling Pump
Incorrect tubing connection to
the sampling pump
Reconnect the tubing to the sampling pump.
Leakage from other
components of the
instrument
Incorrect pump or valve or
tubing connection or bad
sheath in-line filter
Turn off the instrument and contact Agilent
technical support for assistance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals