Running Samples
NovoCyte Quanteon Flow Cytometer Operator's Guide 81
Figure 47. Experiment Control Panel
NOTE
The sample injection probe may perform multiple aspirations for one sampling. Do
not remove or change the test tube until the entire sampling process has
completed.
Note
When Recover Remaining Samples is checked, the remaining unused samples will
be automatically returned to the sample tube. Please refer to Experiment Control in
NovoExpress Software Guide for more details on this function.
During the acquisition, the Cytometer Control > Active Sample Information panel
will show the progress in real-time, as shown in Figure 48.
Figure 48. Active Sample Information
Once the stop condition is reached, sample acquisition will automatically stop.
NOTE
You can also manually stop the sample acquisition by clicking the Stop button in
the Cytometer Control > Experiment Control panel, during the sample acquisition
process.
After sample acquisition is completed, the NovoCyte Quanteon flow cytometer
automatically executes the cleaning process for the fluidics system and the
sample injection probe. This helps to reduce cross-contamination between
different samples. When the cleaning process is completed, repeat the above
procedure for sample NBK.
NOTE
If you want to increase the sampling speed, you can turn off the automatic cleaning
procedure after sampling. To do so, uncheck Rinse after Sampling box in the
Cytometer Control panel from the NovoExpress software as shown in Figure 47.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies NovoCyte Quanteon and is the answer not in the manual?

Agilent Technologies NovoCyte Quanteon Specifications

General IconGeneral
BrandAgilent Technologies
ModelNovoCyte Quanteon
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals