46
Flow Readings
Issue: How can I see ow in dierent units?
Action: From the Main Display, press the button below the ow reading
twice, and then choose Set button eng units. Use the UP/DOWN keys to
move the > cursor to the desired unit, and then press SET.
Issue: The live ow readings won't settle down.
Action: The ow controller is very fast, so it can detect subtle variations
in ow that may go unnoticed by your other ow devices. This
sensitivity can help detect problems with pumps or ow controllers.
You can lessen this sensitivity by decreasing the ow averaging
(press MENU > ADV SETUP > SENSOR SETUP > FLOW AVG
Issue: My ow readings are negative.
Action: Command a zero setpoint to see if the ow returns to 0 after 1.2
seconds. Under conditions of no ow, a negative ow reading can
indicate a poor tare. Ensure that the ow controller has no ow
passing through it, and select TARE FLOW from the Main Display to give
it a fresh tare.
Issue: Does the Alicat work if it is laying down? Will it be accurate?
Action: Yes to both! The ow controller is internally compensated for any
changes in orientation, so you can use it sideways, on its back, or
upside-down. S and QS-series devices should be tared again after
changing their orientation.
Issue: Can I put the Alicat on top of a vibrating device? Will it be accurate?
Action: Yes, and yes! The ow controller is internally compensated for any
changes in orientation, including rapid vibrations. Noise will increase
if the ow meter/controller is vibrating.
Issue: My controller does not agree with another meter I have in line.
Action: Check the STP or NTP settings (MENU > BASIC CONFIG > STP/NTP) to ensure
that your standardized temperature and pressure references match
those of your other ow calibrator. Also check thatyour device's Gas
Select is set to the right gas or mixture.
Issue: My ow readings won't change when ow changes.
Action: If your ow readings won't change regardless of actual ow, your
ow sensor may be damaged. Please contact Alicat to troubleshoot.
Issue: Can I use the Alicat with other gases?
Action: Yes! Your ow controller is designed specically to work with many
dierent gases. Gas Select (MENU > BASIC CONFIG > GAS) includes up
to 130 preloaded gases and gas mixes, or you can dene your own
using COMPOSER. If your desired gas is not listed, please contact
Alicat to ensure compatibility.
Troubleshooting Your Flow Controller

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Alicat MC Series and is the answer not in the manual?

Alicat MC Series Specifications

General IconGeneral
BrandAlicat
ModelMC Series
CategoryController
LanguageEnglish