PN 9000-165-VE, Rev P
Setting Calibration Gas Values
The Calibration Gas Values screen allows you to set and save the calibration
gas values for the sensor you have chosen, before calibrating the analyzer.
These values will be used when calibrating the sensor. However, if the values
are not set here, they can still be set during the calibration process under the
Start Calibration menu.
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
tion Menu.
2. Press the Left or Right arrows to locate the Cal Gas Values menu and press
ENTER/MENU.
3. Enter the values for the gases available for the sensor you have designated
on the HOME screen.
4. Press ENTER/MENU to accept the values you entered.
Figure 2-12.2.
Start Calibration screen.
Figure 2-12.1.
Set Calibration Gas Values
screen (via Cal Gas Value menu).
2-16 | Thermox
®
AMEVision Display User Interface