PN 9000-165-VE, Rev P
All measurements require a two-point calibration: span and zero, so
the zero will always be performed.
Calibrations are only allowed when the sensor is in Normal mode.
Check the Status Bar at the bottom of the HOME screen if you are
unsure of the sensor’s status.
If you do not select a gas to calibrate, a warning displays indicating
that no gas was selected. Press CANCEL to exit the screen.
After a calibration is completed and if you have replaced the analyzer
cell since the last calibration, you will see a yellow screen indicating that
a cell change has been detected and asking if you would like to reset the
Cell Age. If you would, navigate to “Yes” and press ENTER/MENU. If you
would not, navigate to “No” and press ENTER/MENU.
Calibrate the Analyzer
Perform a Manual Calibration
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
tion Menu.
2. Press the Left or Right arrows to locate the Start Calibration menu and
press ENTER/MENU.
3. You are prompted to select the gases to be calibrated. Select the individual
gases or choose ALL.
4. Enter the Recovery Duration time.
5. Press the Right arrow to advance to the next screen.
6. Enter the calibration gas values for the gases you have selected, press the
Up and Down arrows to advance to the elds.
As a default, any gas values that you entered in the Cal Gas Values
menu will automatically be entered on this screen. You can edit these
values and they will be changed in the Cal Gas Values menu as well.
2-18 | Thermox
®
AMEVision Display User Interface