Ametek AMEVision - Optional Blowback Configuration

Ametek AMEVision
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN 9000-165-VE, Rev P
Optional Blowback Conguration
The Blowback Conguration screen can be accessed from the Conguration
menu on the AMEVision controller. For this menu to appear, Blowback must
rst be enabled on the Sensor Conguration screen.
The WDG-V allows you to choose from four Blowback trigger types:
Manual OnlyYou must initiate the Blowback manually from the Utility
Menu. See below.
Timed – Blowback will occur continuously at a predetermined time inter-
val (i.e. every 2 hours). It is recommended that the time interval be at least
1 hour to prevent valve coil overheating.
Flow – Blowback will automatically be triggered if the ow reading from
the ow sensor drops below a user-dened threshold.
Timed and Flow – Blowback will occur at the predetermined time inter-
val unless ow drops below the user-dened threshold in which case the
Blowback will trigger due to low ow.
Low Flow triggered Blowback will only occur once per hour.
Blowback Interval – Time in hours between automatic timed Blowback triggers.
Blowback Recovery – Time in seconds when the oxygen and combustible
readings are frozen after a Blowback has been triggered to allow the system to
return to equilibrium.
Low Flow Threshold – Flow reading in % at which a Blowback is triggered.
Figure 5-1.
Blowback Conguration
screen.
Optional Blow Back Conguration | 5-1