Auto Cal Using the Calibration Start Menu
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
tion Menu.
2. Press the Left or Right arrows to locate the Start Calibration menu and
press ENTER/MENU.
3. Navigate to the drop down list and press ENTER/MENU.
Select “Automatic.”
Figure 2-16.
Select calibration type.
4. Press the Right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Cal Gas Duration in seconds.
7. Enter the time for the Recovery Duration in minutes.
8. Press the Right arrow to continue.
This is the Calibration Gas Values screen. If you have already entered
your calibration gas values in the Cal Gas Menu, they will be visible in
the elds. Any changes that you make to these values will be reected
in the Cal Gas Menu as well.
9. When you are satised with your values, press ENTER/MENU
to start the automatic calibration. The screen displays the Gas Value,
Cal Gas and Time Remaining. To abort the calibration, navigate to the
ABORT button and press ENTER/MENU.
If O
2
is selected for calibration, it will be the rst gas calibrated and
the zero gas will be calibrated next, then combustibles and methane.
If combustibles is the rst gas selected for calibration, zero gas will be
calibrated rst, then combustibles and methane.