Perform an Automatic Calibration
The functions for automatic calibration will only be visible if you have
an RCU and have congured your sensor for Automatic Calibration
in the Sensor Conguration menu.
There are two ways to perform an automatic calibration:
• Dene the automatic calibration time and date in the Auto-Calibration
Conguration (Auto Cal Cong) menu. This allows the system to perform
calibrations at scheduled times and there is no further user interaction.
• Use the Start Calibration (or Set Cal Gas) menu to dene the gases and
values and then use the Start Calibration menu to perform an immedi-
ate calibration. This requires you to program the Start Calibration menu
and physically press ENTER/MENU to start the calibration. The real-time
sequence of calibration events is displayed on a series of screens which
require user before each step of the calibration continues.
Conguring the Auto Cal Using the Auto-Cal Cong Menu
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
tion Menu.
2. Press the Left or Right arrows to locate the Auto Cal Cong menu and
press ENTER/MENU.
3. Press the Left and Right arrows to navigate to the next screen.
Figure 2-15.
Auto-Calibration
Conguration screen.