Controller/Interface | 2-21
PN 9000-165-VE, Rev P
Perform an Automatic Calibration
The functions for automatic calibration will only be visible if you have
an RCU and have congured your sensor for Automatic Calibration
in the Sensor Conguration menu.
There are two ways to perform an automatic calibration:
Dene the automatic calibration time and date in the Auto-Calibration
Conguration (Auto Cal Cong) menu. This allows the system to perform
calibrations at scheduled times and there is no further user interaction.
Use the Start Calibration (or Set Cal Gas) menu to dene the gases and
values and then use the Start Calibration menu to perform an immedi-
ate calibration. This requires you to program the Start Calibration menu
and physically press ENTER/MENU to start the calibration. The real-time
sequence of calibration events is displayed on a series of screens which
require user before each step of the calibration continues.
Conguring the Auto Cal Using the Auto-Cal Cong Menu
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
tion Menu.
2. Press the Left or Right arrows to locate the Auto Cal Cong menu and
press ENTER/MENU.
3. Press the Left and Right arrows to navigate to the next screen.
Figure 2-15.
Auto-Calibration
Conguration screen.