14 User Guide EN Dansensor® LeakPointer 3/3
+
P/N 380064-F
06/2021
3. The measurement result is converted to a fictive hole size (in m) corresponding to a
standard round hole in a 0.13 mm thick foil (if device is calibrated using the optional “Leak
Simulation Kit” accessory - see
"Options/Accessories" on page 91).
This does not mean that a hole of this size exists in the package, instead there could be 3
holes in the package, and together they correspond to the calculated standard hole size.
4. When the measurement is complete, the vacuum is released from within the chamber and
the chamber
opens.