Electrical specifications
Power supply LeakPointer 3: 100 - 264 VAC + PE, 47- 63 Hz
Fuse 5x20 mm, 2 AT
LeakPointer 3
+
, 115V: 103 - 127 VAC + PE, 57- 63 Hz
Fuse 5x20 mm, 10 AT
LeakPointer 3
+
, 230V: 207 - 253 VAC + PE, 47- 53 Hz
Fuse 5x20 mm, 6.3 AT
Power consumption LeakPointer 3: Max. 50 Watt
LeakPointer 3
+
, 115V: Max. 900 Watt
LeakPointer 3
+
, 230V: Max. 900 Watt
Connectivity
USB 2 x USB (2.0 - Type A), max. current 500 mA
For connection of bar-code scanner, printer and memory stick
Network/LAN RJ-45 Ethernet 10/100 Mbit/s, DHCP client or fixed IP
Compressed air supply
(LeakPointer 3 only)
Hose dimensions Ø6/4 mm
1
Air supply pressure 5.5 ± 0.5 Bar (80 ± 7 psi)
Air consumption Approx. 33 l/measurement, max. 162 l/min
Air quality Dry, clean and free from oil
Compliant with DIN ISO 8573-1:2010 [6:4:4]
2
1
Ensure that the used hose type is appropriate for the required pressure
2
See "Filter recommendations" on page 17
CO
2
Sensor
Sensor type Non Dispersive Infra Red (NDIR), single beam
Sensor life Approx. 3 years