EasyManuals Logo

Ametek Mocon Dansensor LeakPointer 3 User Manual

Ametek Mocon Dansensor LeakPointer 3
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Dansensor® LeakPointer 3/3
+
EN User Guide 75
P/N 380064-F
06/2021
Information
The Information screen holds general information about the device.
Diagnostics
The Diagnostics screen holds various internal device parameters.
The parameters can only be read and not changed.
Serial no. Shows the device’s serial number.
Firmware Version Shows the firmware version currently installed on the device.
Number of tests Shows the total number of tests performed.
Next calibration Shows the number of days left until the device should be calibrated.
Chamber switch Shows if the chamber is open or not.
Can be used to check if the magnets in the sealing are placed
correctly.
Temperature Shows the temperature measured on the PCB in the device.
This will always be slightly higher than the surrounding
temperature.
CO2 Shows the current CO
2
concentration in the surrounding
atmosphere.
Chamber Fan Speed Shows the current speed of the fan inside the chamber.
Chamber pressure (U3) Shows the pressure difference between the chamber and the
surrounding atmosphere.
This should be close to 0 when the chamber is open.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Mocon Dansensor LeakPointer 3 and is the answer not in the manual?

Ametek Mocon Dansensor LeakPointer 3 Specifications

General IconGeneral
BrandAmetek
ModelMocon Dansensor LeakPointer 3
CategoryTest Equipment
LanguageEnglish

Related product manuals