Measuring
CO
2
levels
High or unstable background CO
2
levels can affect the function of the device as well as the
measurements made in the device.
The device responds as follows in the situations described below:
If the CO
2
level in the chamber exceeds a certain value (default = 7500 ppm) before a
measurement is started or if the
CO
2
value in the chamber exceeds a certain value (default
= 9500 ppm) during a current measurement, the process is interrupted and the following
error message appears:
In these cases, it is recommended to open the chamber and wait approx. 30 sec. before
trying again.
In some cases it may even be necessary to ventilate the room where the device is located,
at the same time.
If the increase in the CO
2
level in the chamber exceeds a certain value (default = 5000 ppm)
during a current measurement, the process is interrupted and on the measurement screen
the result is indicated as a "LEAK" of either 999 µm or 99.9 ppm/s depending on the
selected measurement method.
Selecting a product for measuring
1. In the measuring screen....
... tap the icon in the top left corner. This will bring up the Select Product screen.
(This screen can also be reached by tapping the icon in the bottom right corner, and
then selecting Products -> Select (as active).