16 User Guide EN Dansensor® LeakPointer 3/3
+
P/N 380064-F
06/2021
Connections
Main power supply cable
CAUTION! The main power supply cable must be prepared by authorized
personnel.
The enclosed main power supply cable is provided “open-ended” (i.e. without mains plug).
Install a mains plug that meets the local requirements.
WARNING! The ground wire (green/yellow) must always be connected to the local
safety ground system in accordance with the local requirements.
Alternatively a standard power cable (C14 acc. to IEC 60320-1, UL 498 or CSA C22.2 no. 42)
with a local mains plug can be used.
Minimum rating must be in accordance with the requirements - see "Electrical specifications"
on page 88.
Connect the cable between a suitable power outlet and the device’s “POWER SUPPLY”
connector. Make sure that the yellow locking mechanism locks the cable in the connector
(pull the cable gently to check).