44 User Guide EN Dansensor® LeakPointer 3/3
+
P/N 380064-F
06/2021
Setting up products for “Multi Package” testing
The setup procedure for Multi Package products includes the following:
Create product - see "New" on page 58.
Determination of vacuum setting
1. Prepare the test package for DeltaP measuring as described in "Determination of vacuum
setting" on page 31.
NOTE! Make sure that neither the needle nor the hose is hit by the chamber
when the chamber is closed. We recommend to place the needle through the
side of the package.
2. Select Workflow -> Products -> Edit and then select the product you have
created for the specific package type.
3. Set Mode to Multi Package.
4. Tap the yellow icon to go to the Create Test Vacuum screen.
5. Set the Chamber vacuum (delta) to 300 mbar, then close the chamber to start the
measurement.
6. The chamber will be evacuated to the set value and the measured values are continuously
displayed on the screen.
NOTE! If the currently measured “CO2” value increases when vacuum is applied,
then the package is most likely leaking and you should thus use another
package to perform the test.