66 User Guide EN Dansensor® LeakPointer 3/3
+
P/N 380064-F
06/2021
Access levels
The various access levels and the associated rights are described below:
NOTE! A higher level of access always has all rights from lower access levels.
Supervisor login
At start-up when user login is required and you are logged in either by selecting a user from
the
User Login list or by entering a valid User ID whatever is required, even users created
with Supervisor access will also only have Operator access at this point.
To obtain Supervisor rights, you must log in again via Main Menu -> Access -> User
Login with a valid User ID.
Standard Test
Operator
Basic user.
Can only perform tests on workflows/products marked as
Standard Test.
If Use State Mark is enabled in Settings -> Device the access
is limited to Standard Test products/workflows marked with
State = Current.
Cannot change any settings.
Operator Normal user.
Can perform tests using all products/workflows with
State = Current.
Cannot change any settings.
Supervisor Daily administrator of products/workflows and standard users.
Has access to all products/workflows.
Can create products/workflows and edit custom fields.
Can create Standard Test Operator and Operator users, but not
other Supervisor users.
Can access the System Test and Create Test Vacuum functions
from the Service -> Test & Adjustment menu.
Can export measurement logs and error/event logs.
Cannot change settings in the Settings menu.
Administrator Device administrator.
Can change settings in the Settings menu (except for the
Service menu).
Can create Supervisor users.
Can remove access to Service login.
Can export and import all data.
Service Service Technician.
Can access all settings in the Settings -> Service menu.
Can access all functions in the Service -> Test & Adjustment
menu.