Dansensor® LeakPointer 3/3
+
EN User Guide 83
P/N 380064-F
06/2021
Network Logging
The Network Logging screen holds the parameters for the network logging function.
If enabled, the function sends a data string through a LAN connection for each measurement.
NOTE! Always consult your network administrator before making network
settings, as incorrect settings can result in reduced or no network activity.
A TCP/IP LAN server has to be implemented to receive data. The server will listen to the port as
specified above.
The protocol is an array of semicolon separated values:
"Device Serial No; Date; Time; User Name; Product Name; Product ID; Mode; Standard Test;
State; Vacuum (delta); Test time; Leak limit; Hole limit; Leak gain factor; Hole size (micron); CO2
increase /ppm/s); Leak; Marked state; Error nr; Days till calibration; Invalid measurement
comment; Custom field 1; Custom field 1 reply; Custom field 2; Custom field 2 reply; Custom
field 3; Custom field 3 reply; Custom field 4; Custom field 4 reply;"
Network Logging Select whether or not network logging is required.
Disabled = No network logging
Enabling Network Logging also enables the following
parameters:
Server IP Set up an IP-address to be used for
collecting log data for each measurement
via LAN.
This requires for setting up of a Server Port
number as well.
Server Port See above.
Acknowledge Byte sent by the server to the device after
each received log entry.
Setting value to 0 will disable the function.