Specications | 2-3
Sensor
Operating Range
0.1 PPM O
2
to 100% O
2
Accuracy
Percent: ±1.0% of reading or .02% O
2
absolute, whichever is greater.
PPM: ± 2% of reading or 0.5 PPM O
2
absolute, whichever is greater.
Response Time
< 5 seconds at 0.6 L/min (1.3 scfh) over one decade.
Repeatability
Percent: ±0.5% of reading or 0.1% O
2
absolute, whichever is greater.
PPM: ± 0.5% of reading or 0.1 PPM O
2
absolute, whichever is greater.
Ambient Temperature
-18° C to 50° C (0°F to 122°F)
Max. Inlet Temperature
160° F (71° C)
Sample Flow
0.1 to 1.0 L/min (0.2 to 2.1 scfh).
Environment
IEC Installation Category II
IEC Pollution Degree 2
Humidity: 5 to 90%, non-condensing
Altitude: 2000 meters