2-4 | Thermox TM2000 Oxygen Analyzer
Enclosure
Indoor/Outdoor, Type 3R.
Power Requirements
115 VAC, ± 10%, 50/60 Hz, 288 VA max. (230 VAC optional)
Calibration Gas Requirements
0.1 to 1.0 L/min (0.2 to 2.1 scfh) (0.6 L/min recommended).
Zero Gas: From 0.1 PPM to 10% O
2
, balance N
2
Span Gas: Minimum one decade above zero gas
(10 times greater).
System Compliance
EMC Directive 2004/108/EC
Low Voltage Directive 73-23/EEC
Remote Calibration Unit (RCU):
PPM RCU
Enclosure:
Indoor/Outdoor; UL Type 4X (NEMA 4X) (IP56)
Envoironment:
Ambient Temperature: -18° C to 50° C (0° F to 122° F)
Humidity: 0 to 90%, non-condensing
Max Altitude: 2000 Meters
IEC Installation Category: II
IEC Pollution Degree 2