4-18 | Thermox TM2000 Oxygen Analyzer
Remote Switch Conguration
If initiating an automatic calibration or verification using a digital input,
this menu option allows you to define whether the input should trigger a
calibration or a verify (Remote Switch Config).
See the Installation chapter for help on connecting wiring between the
digital input and the TM2000 control unit to activate a digital input signal.
Conguring the Remote Switch
To define whether the digital input should trigger an automatic calibration
or an automatic verify, do the following:
1. Select Remote Switch Config from the Setup Key.
2. Select Calibrate or Verify.