4-26 | Thermox TM2000 Oxygen Analyzer
You will be prompted to enter an output filtering value:
[100=FAST 1=SLOW
Fltr = XX New #?
where XX is the current filter value. Note that 100 equals no filtering, and
1 equals the highest possible filter value.
2. Type a number between 1 and 100, then press the Enter Key. Note that
you should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
The default value is 100 (no filtering).
Select Function Option
The Select Function menu option allows you to select what type of infor-
mation the analog output port should track. Choices are as follows:
Oxygen
Cell temperature
Thermocouple millivolts (T/C mv)
Cell millivolts (cell mv)
Selecting the function
1. Choose Select Function from the Analog Range menu after selecting
an analog output port.
2. Select the function you want the selected analog output port to track
(see bullets in Description section above).
The Series 2000 control unit then prompts you with the Set Current
Range menu option where you choose the range of readings that the 0
to 20 mA or 4 to 20 mA analog output represents (see the “Set Current
Range” section in this chapter for details).