EasyManua.ls Logo

Anritsu Spectrum Master MS2721B - 3-2 Measurement Setup

Anritsu Spectrum Master MS2721B
137 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-2 Measurement Setup Chapter 3 — Quick Start Guide
3-2 PN: 10580-00175 Rev. N MS272xB + MS8911B UG
3-2 Measurement Setup
Connecting the Input Source
Connect the input signal or antenna to the appropriate test connector on the top of the
instrument. For connector descriptions, refer to Figure 2-10 on page 2-13.
Editing and Entering Values
Parameter values that are ready for editing are displayed in red on the submenu key.
After changing the value, press Enter to set the new value.
Some submenu keys have toggled parameter values (On / Off, Low / High). On these
submenu keys, the current value is underlined. Press the submenu key to toggle the
value.
Use the arrow keys, numeric keypad, or rotary knob to change submenu key values,
to select list box options, or to enter filenames.
Figure 3-1. Spectrogram Example

Table of Contents