A-2 Test Records for Spectrum Analyzer Verification Appendix A — Test Records
A-2 PN: 10580-00279 Rev. E MS272xC MM
A-2 Test Records for Spectrum Analyzer Verification
MS272__C Firmware Rev: ______________ Operator: ____________________ Date: ____________
Serial Number: _______________ Options: __________________________________________________
Table A-1. Spectrum Analyzer Frequency Accuracy
Frequency Measured Value Error Specification
1GHz GHz Hz ± 1300Hz
7 GHz (MS2722C only) GHz Hz ± 9100 Hz
12 GHz(MS2723C only) GHz Hz ± 15600 Hz
18 GHz(MS2724C only) GHz Hz ± 23400 Hz
30 GHz (MS2725C only) GHz Hz ± 39000 Hz
40 GHz (MS2726C only) GHz Hz ± 52000 Hz
Table A-2. Spectrum Analyzer 1.0 GHz SSB Phase Noise Verification
Frequency Offset Measured Value Specification
10 kHz dBc/Hz –100 dBc/Hz
100 kHz dBc/Hz –102 dBc/Hz
1000 kHz dBc/Hz –107 dBc/Hz
10 MHz dBc/Hz –120 dBc/Hz
Table A-3. Spectrum Analyzer Second Harmonic Distortion
Frequency Measured Value Second Harmonic Distortion Specification
50.1 MHz dBm
100.2 MHz dBm dBc –54 dBc max