2-10 Third Order Intercept (TOI) Chapter 2 — Spectrum Analyzer Verification
2-22 PN: 10580-00279 Rev. E MS272xC MM
2-10 Third Order Intercept (TOI)
The following test is used to verify the Displayed Average Noise Level (DANL) of the spectrum analyzer
systems in the MS272xC Spectrum Master. This test is performed using the RMS detection mode.
The following test verifies the Third Order Intercept point (also known as TOI or IP3) of the spectrum analyzer
in Spectrum Master Models MS2722C, MS2723C, MS2724C, MS2725C, and MS2726C.
2400 MHz TOI Test
Equipment Required for MS2722C, MS2723C, and MS2724C
Anritsu MG3692x Synthesized Signal Generator
Frequency Reference Symmetricom Rubisource T&M
Adapter, Anritsu Model 34NN50A
Spectrum Master: MS272xC
Power Meter, Anritsu Model ML2438A
Signal Generator, Agilent Model 8648D
Power Splitter, Aeroflex/Weinschel Model 1870A
Power Sensor, Anritsu Model MA2442D (10 MHz to 18 GHz, –67 dB to +20 dB)
Fixed Attenuator, Aeroflex/Weinschel Model 442
Fixed Attenuator, Aeroflex/Weinschel Model 446
Fixed Attenuator, Aeroflex/Weinschel Model 4420
RF Coaxial Cable, N(m) to N(f), Anritsu Model 15NNF50-1.5C (3)
Equipment Required for MS2725C and MS2726C
All of the above, plus
Anritsu 71693-R Ruggedized K(f) to N(f) Adapter