Chapter 2 — Spectrum Analyzer Verification 2-2 Frequency Accuracy Verification
MS272xC MM PN: 10580-00279 Rev. E 2-3
18. Set the MG3692x or MG3695C frequency to:
7 GHz for MS2722C
12 GHz for MS2723C
18 GHz for MS2724C
30 GHz for MS2725C
40 GHz for MS2726C
19. Set the MS272xC center frequency to:
7 GHz for MS2722C
12 GHz for MS2723C
18 GHz for MS2724C
30 GHz for MS2725C
40 GHz for MS2726C
20. Press the Marker main menu key, and select the Peak Search submenu key.
21. Verify that the marker frequency is:
7 GHz ± 2.1 kHz (± 0.3 ppm) for MS2722C
12 GHz ± 3.6 kHz (± 0.3 ppm) for MS2723C
18 GHz ± 5.4 kHz (± 0.3 ppm) for MS2724C
30 GHz ± 9.0 kHz (± 0.3 ppm) for MS2725C
40 GHz ± 12.0 kHz (± 0.3 ppm) for MS2726C
22. Record the results in the test records. Use Table A-1.
Note
If the unit fails the Spectrum Analyzer Frequency Accuracy test, replace the SPA PCB assembly.
Refer to “Spectrum Analyzer Module Assembly Replacement ” on page 5-5.