Anyview A8 - 13.4 Setting CO2

Default Icon
305 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient monitor User’s manual
Chapter 13-7
13.4 Setting CO2
Select the CO
2
parameter area to enter CO
2
parameter setting menu.
Setting Unit
Select 【 Unit 】 in the CO
2
parameter setting menu, and the options are
【mmHg】,【kPa】or【%】.
Setting Apnea Alarm Time
For mainstream CO
2
and microstream CO
2
module, select【Apnea Alm】in
the CO2 parameter setting menu, you can select the apnea time as required in
the options. The monitor indicates an alarm when a pre-adjusted time has
elapsed since the last detected breath.
Setting Expiring Cycle
The value of EtCO
2
and FiCO
2
in the CO
2
parameter area will be refreshed
in real-time. As for the mainstream CO
2
module and microstream CO
2
module,
you can set the method to calculate EtCO
2
and FiCO
2.
The【Et Cycle】can be set in the CO
2
parameter setting menu:
【One Breath】:calculate the EtCO
2
and FiCO
2
by every respiratory wave.
【 10s 】 or 【 20s 】: a time interval, during which the maximum CO
2
concentration is EtCO
2
, and the minimum CO
2
concentration is FiCO
2.
Setting CO
2
Corrections
Temperature, water vapor in the patient’s breath, barometric pressure, and
the proportions of O
2
, N
2
O and Helium in the mixture all influence CO
2
absorption.
For mainstream CO
2
module and microstream CO
2
module, select【Set CO
2
Calibra. 】 in the CO
2
parameter setting menu, set these options as required
before zeroing:
——【Gas Temp】:Set the temperature of gas.
——【Barometric】:Set the atmospheric pressure.
——
【
Calibrate Gas
】
:Select the gas type of zeroing, the options are
【
Air
】
and【N
2
】.
—— 【 O
2
Compens 】 :Select the concentration of oxygen. It can be set to a